RI Title
RI logo

LIMITED AVAILABILITY

This item will not be restocked when depleted and should be avoided for new designs.
This product entered limited availablity on 06/20/2018. View RIKs with Limited Availability for a complete list.

Suggested Replacement: RI8589C

Expiration Reason: The 'C' upgrades include better accuracy at high voltages and the addition of an 'HV4' negative supply aimed at GAN devices.

Aditional replacements may be found by searching for RI8589A (change the REV letter) or High Power Supply/Pulser TIM.

High Power Supply/Pulser TIM

Part Number: RI8589A5

Please visit the Instruments page for the latest Briefs.

Test Instrument Modules Browse Instruments »

FET Power Pulse TIM




PRELIMINARY SPECS (Verification pending)

1.0 Gate Bias
V max 60v
V min -10v
V set resolution 5mv
V measure ranges 100v, 10v, 1v, 100mv.
V measure accuracy .1%
V settling time 10uS
R out 100 ohms in high current range
1000 ohms in low current range
V set accuracy .2% of set voltage
I max 10ma ( could be 50 ma)
I measure ranges 100mA,10mA,1mA,100uA,10uA,1uA
I measure accuracy .2%
Measure Rate 80KspS
Output switch
blocking 200 volts
leakage 100 pA

protected against shorts
no current set capability or adjustable limits

2.0 Drain Bias
V max 175v (two ranges 90v and 175v)
V min 0v
V set resolution 20mv
V measure ranges 200v, 20v, 2v, 200mv.
V measure accuracy .2%
V settling time 10uS
R out 100 ohms in high current range
1000 ohms in low current range
V set accuracy .2% of set voltage
I max 10mA
I measure ranges 10mA,1mA,100uA,10uA,1uA
I measure accuracy .2%
Measure Rate 80KspS
Output switch
blocking 200 volts
leakage 100 pA

protected against shorts
no current set capability or adjustable limits

3.0 Drain Pulsing
V max 50v
V min 8v
V set resolution 20mv
V set accuracy .2% of set voltage
V measure ranges 200v, 20v, 2v, 200mv.
V measure accuracy .2%

I max 20A
I measure ranges 20A,2A,1A,100mA,10mA
I measure accuracy .3%

I limit range 1 to 20A
I limit resolution 100mA

Measure Rate 80KspS
Output switch
blocking 500 volts
leakage 100 pA

Pulse width min 10 uS
Pulse width max > 100mS set by energy in pulse
Pulse rep rate 10 % duty cycle min based on energy in pulse

Protected for output shorts.
Has settable current limiter.

4.0 Drain reverse diode test
R out 2.5 ohms
I max 10A
V max - 50v
V min - 5v

5.0 Drain reverse diode temperature testing
I test -10mA
Delay to first measure 10uS

6.0 Relay driver
4 lines to control external high current relays

Related Documents:

User account required Create new account  
RI8589A4_ButtonGuide_BulkPwr_FET_TIM_Dec2015revE.pdfRI8589A4_ButtonGuide_BulkPwr_FET_TIM_Dec2015revE.pdfRI8589A4_HypertronicsInterface_Dec2015revE.pdfRI8589A4_HypertronicsInterface_Dec2015revE.pdf

RI8589A5_BlockDia_BulkPwr_FET_TIM_May2018revG.pdfRI8589A5_BlockDia_BulkPwr_FET_TIM_May2018revG.pdf