Cassini16 can be diagnosed and calibrated with the Saturn Guru application. Follow the steps below to perform a full system Cal. First run Diagnostics to collect the "before" data and report any failures to RI Support for review prior to Calibration. Once the TIMs are confirmed functional, perform the Calibration. After the calibration procedure is completed, run Diagnostics a final time to collect the "after" data.



Related Documents:
Requirements:
  • Cassini Calibration Kit (Power Meter, DMM, etc.)
  • Cassini Diag Kit (Configuration Specific) with Diag/Cal Interface Plate, Adapters and Cables
  • Cassini Fixture

To Diagnose Fixture with Saturn on Win10:
  1. Power on the Win10 EPC and Cassini 16 Infrastructure
  2. Log on to Windows using the shared Roos Admin username and password.
  3. Saturn will launch automatically by Guru. (Edit with Guru Agent Editor) Otherwise, Choose Log on from Guru and choose User Apps or Apps and select Saturn Win10 to launch the application.
  4. Wait to proceed until "Tester Loaded." is shown. Start a 20 minute timer to allow instruments to warm up.
  5. If the Fixture is latched, then it will be added automatically at startup. If it is not latched, latch the Fixture if it is not yet attached. From Saturn app, choose apps > cassini, select equipment and choose Launch app form the right mouse button menu.
  6. From the equip pool window, choose refresh from the right mouse button menu. From Saturn app, choose apps > browse tester and confirm the Fixture is similar to "ACME Cart"
  7. From Saturn app, choose apps > cassini, select cal exec and choose Launch App from the right mouse button menu.
  8. Choose File > Diag List and select the appropriate Diag list (likely "NG PA Cart Rev A").
  9. Select each service plan (CTL+A or CTL+left mouse button click) and choose choose Run Selected from the right mouse button menu. Remember to choose Actions > Results and Data > Save Csv to record the changes for future analysis.
  10. If any FAIL results are found, follow the steps below to Send Diagnose Results to [email protected]. If all results PASS, proceed To Calibrate NG Cart with Saturn on Win10 below.
  11. Close the cal exec window.
To Calibrate Fixture with Saturn on Win10:
  1. Follow the steps to Diagnose Cassini 16 with Saturn on Win10 above.
  2. Choose Apps > Cassini, select Fixture Cal Exec and choose Launch App from the right mouse button menu.
  3. Choose File > Load Cal Kit and choose the appropriate entry. RiCalKit naming convention sets the date as the last valid cal date and should be checked if in the past. The Cal Kit contains the Open Coefficients and Noise Source ENR table and should be valid. (See Updating Coefficients of Standards used in a Cal Kit Definition on Cassini https://roos.com/docs/RBEH-8K525N?Open)
  4. Choose between 2 and 6 Cal and Validate pairs by holding CTL and clicking on each entry.
  5. Choose Run Selected from the right mouse button menu.
  6. Follow the user prompts. These often require connecting PKZ#12 adapters to either a Power Sensor, SMA cable, and OSL standards. (See Figure 1 and Figure 2)
    IMPORTANT: ALWAYS follow good RF connector hygiene and use a SMA Torque wrench when tightening and avoid rotating the inner cable, ALWAYS rotate the outer nut.
  7. Repeat until the list is completed. Cal data is ONLY saved after every successful validate. If a validate fails, repeat all selected Cal plans.
  8. Follow the steps to Diagnose Fixture with Saturn above (steps 8-10) to document the "after" values.

To Diagnose Cassini 16 with Saturn on Win10:
  1. Power on the Win10 EPC and Cassini 16 Infrastructure
  2. Log on to Windows using the shared Roos Admin username and password.
  3. Saturn will launch automatically by Guru. (Edit with Guru Agent Editor) Otherwise, Choose Log on from Guru and choose User Apps or Apps and select Saturn Win10 to launch the application.
  4. Wait to proceed until "Tester Loaded." is shown. Start a 20 minute timer to allow instruments to warm up.
  5. Unlatch the Fixture if it is attached and Latch the Diag/Cal Interface plate (see Diagnostic Kit).
  6. From Saturn app, choose apps > cassini, select equipment and choose Launch app form the right mouse button menu.
  7. From the equip pool window, choose refresh from the right mouse button menu.
  8. From Saturn app, choose apps > browse tester and confirm the Fixture is similar to "ACME bottomplate ##"
  9. From Saturn app, choose apps > cassini, select cal exec and choose Launch App from the right mouse button menu.
  10. Choose File > Diag List and select the appropriate Diag list (likely "Acme Rev A").
  11. Select ALL service plans (CTL+A or CTL+left mouse button click) and choose choose Run Selected from the right mouse button menu.
  12. If any FAIL results are found, follow the steps below to Send Diagnose Results to [email protected]. If all results PASS, proceed To Calibrate Cassini 16 with Saturn on Win10 below.
  13. Close the cal exec window.

To Calibrate Cassini 16 with Saturn on Win10:
  1. Follow the steps to Diagnose Cassini 16 with Saturn on Win10 above.
  2. Choose Apps > Cassini, select Cal Exec and choose Launch App from the right mouse button menu.
  3. Choose File > Load Cal Kit and choose the appropriate entry. RiCalKit naming convention sets the date as the last valid cal date and should be checked if in the past. The Cal Kit contains the Open Coefficients and Noise Source ENR table and should be valid. (See Updating Coefficients of Standards used in a Cal Kit Definition on Cassini https://roos.com/docs/RBEH-8K525N?Open)
  4. Choose File > Calibration List and select the appropriate entry (likely "Acme Cal List Rev A")
  5. Choose between 2 and 6 Cal and Validate pairs by holding CTL and clicking on each entry.
  6. Choose Run Selected from the right mouse button menu.
  7. Follow the user prompts. These often require connecting PKZ#12 adapters to either a Power Sensor, SMA cable, and OSL standards. (See Figure 1 and Figure 2)
    IMPORTANT: ALWAYS follow good RF connector hygiene and use a SMA Torque wrench when tightening and avoid rotating the inner cable, ALWAYS rotate the outer nut.
  8. Repeat until the list is completed. Cal data is ONLY saved after every successful validate. If a validate fails, repeat all selected Cal plans.
  9. Follow the steps to Diagnose Cassini 16 above (steps 8-10) to document the "after" values.
  10. Record the model and serial numbers of the instruments along with last cal and next cal dates and generate a Calibration Certificate that includes the next scheduled Cal Due date in 12 months.
    (See Exporting Cassini Calibration Data for Certification or Analysis https://roos.com/docs/RBEH-C6DKWY?Open)

To View the Worksheet from Cal Exec and Save Csv:
  1. Open the Cal Exec results worksheet, click on Actions > Results, to open the worksheet window.
  2. (Optional) Choose Data > Save Csv to convert the data in the worksheet to a .csv file in the %USERPROFILE%\RiApps\data directory. RITdb.datalog is always saved to Guru.
    NOTE: If the .csv file does not appear, follow steps
    "To Setup Python3 For Saving Results as CSV" on the Database 'Product Docs', View 'All Documents', Document 'Saturn Win10 User Guide'Saturn Win10 User Guide. (Administrator access required)


To Send Diagnose Results to Roos Instruments' Support:
  1. After collecting Diag results from the Cal exec, choose Apps > Guru Browser from Guru.
  2. Choose RITdb.datalog from the Key1: ri.sys.ObjClass dropdown menu.
  3. Under Key 2, choose ri.sys.Type and select Validate.
  4. Select all the recent results in the object table and choose Export > As Guru File from the right mouse button menu.
  5. Choose OK to close the Comments window, choose a location and enter a file name and choose Save.
  6. Attach the file to an email to [email protected] and include the Tester's serial number, reason that Diags were run, and a brief description of how to recreate the issue.


Figure 1: Example PMeter Connections
Image.jpegImage.jpeg


Figure 2: Power Sensor N to SMA Adapters (Male - Female)