A new test was added to confirm the functionality of the RI8589 FET Pulser temperature sense feature to was added to service plans RI8589A Current Measure Calibration, RI8589A FET Pulser Verify, and RI8589A FET Pulser Verify FetPulser2 (for configurations with two TIMs).



IMPORTANT: Does not change Calibration or Diag procedure. Does not affect Calibration data.

The 10mA constant current source was not validated in the previous versions of these plans dated prior to 2018-04-04. In the RI8589A Current Measure Calibration test plan, no new calibration data is stored, it is only used to confirm functionality during the calibration process. The Verify compares the measured value to expected values and is accessed by selecting the TIM from the Configuration window and choosing Calibration > Diagnose to open the Diag Exec. The plans are already be included in the Cal List that defines the order of the service plans as they appear in the Cal Exec and Diag Exec.

NOTE: This also includes minor improvements added in 2017 to reduce the likelihood of a false positive failure on the first point of the 37V Pulse test that could measure prematurely.

Status: General Availability, Does Not Affect Measurements, Improves Failure Diagnosis

Action Required
1) To update these Cal Plans, import the attached Ri8589A_FetPulserCalDiagPlans_Update20180501.gzp file below.

Ri8589A_FetPulserCalDiagPlans_Update20180501.gzpRi8589A_FetPulserCalDiagPlans_Update20180501.gzp

2) Confirm RI8589A Current Measure Calibrationexist in the Cal Exec selecting Tester > Calibrate and and RI8589A FET Pulser Verify (FetPulser2) exists in the Diag Exec by selecting Tester > Diagnose from the Configuration window (System > Tester). Please contact [email protected] and request the RiCalList object to be modified to include these testplans if they are not displayed.

Example Output
Test called "TemperatureSense" will Pass with Max value limit of -5.

Example results:

Test Result Data
------- ------- -------
TemperatureSense Pass -9.34322342  


Test Result Data
------- ------- -------
TemperatureSense Fail -1.34322342  

Affected TIMs
15 TIMs have been assembled and shipped prior to 01/26/2018 that would benefit from these updated Cal/Diag plans.
TIM Model Numbers: RI8589A2, RI8589A3, RI8589A4, RI8589A5
Model - Serial Numbers:
A2 - RRE2RK0A
A3 - RRRC220A, RRWCWK0A
A4 - RPMV0T0A, RRGCCC0A, RRWCYE0A, RS37U90A, RS37UK0A
A5 - RTF0B60A, RTGZGM0A, RTZTMM0A, RU4S380A, RU64UZ0A, RU79F30A, RU9SNG0A