Improved physical interfaces for easier integration than RIK0017A. Adds Scan and 16 MegaVector memory depth. Use RIK0148A Digital Coax cables ONLY. DO NOT USE DC JUMPER WIRES.

Provides High Speed (Up to 80 Mbits/sec.) Digital control signals to the DUT from the fixture.




Top Board: = Y0008NA1


High Speed Scan Module Kit (1 Module)

2 Sets of 4 I/O Pins: DB 1 to 4 and DB 5 to 8
Von and Voff Set Range: 0 to 4V
Vcompare: 0 to 4V
Serial Control Supports: Data Read, Data Write, Clock, Strobe
Serial Write: Up to 80 Mbits/sec.
Scan Support: Yes
Parametric Measure: Yes


Y0008NA1 HS Digital Scan Pin Out

Pin#
Function
Note:
Connector Type
P1
Configurable
Data, Clock, Strobe, or Static
Header Pin
P2
Configurable
Data, Clock, Strobe, or Static
Header Pin
P3
Configurable
Data, Clock, Strobe, or Static
Header Pin
P4
Read A
Header Pin
P5
Read B
Header Pin
P6
Configurable
Data, Clock, Strobe, or Static
Header Pin
P7
Configurable
Data, Clock, Strobe, or Static
Header Pin
P8
Configurable
Data, Clock, Strobe, or Static
Header Pin
P9-16
GND
H4-1
GND
Header Pin
H4-2
MUX I/O
Header Pin


Configuration Window Settings:
Module = Y2000H1A
Inst#: RI74315C
Program = Y18NA11A

Vlo = 0
Vneg = 2
Vector Freq = 80000000
Vhi = 0
Vcmp = 4
Clock Freq = 20000000
DB1 = off

Scan Location B = blank
Scan Location = blank

NOTES:
  1. Module will only work in M1 or M5 positions of the carrier board.
  2. Keep 40 pin ribbon cables from fixture bottom to carrier less than 8".






Configuring the Carrier:

RIK0127A_HSDigConfiguration_RI7100.pdfRIK0127A_HSDigConfiguration_RI7100.odp

Scan Requirements:
tbd.